Assessment of the Pore Diameter Measured in Gas-Liquid Displacement Porometry: Effect of the Capillary Constant

Ängeslevä, Marina; Salmimies, Riina; Rideal, Graham; Häkkinen, Antti (2022-11-16)

Katso/Avaa
angesleva_et_al_assessment_of_the_pore_diameter_measured_aam.pdf (1.261Mb)
Lataukset: 


Post-print / Final draft

Ängeslevä, Marina
Salmimies, Riina
Rideal, Graham
Häkkinen, Antti
16.11.2022

Journal of Porous Media

26

1

31-49

Begell House

School of Engineering Science

Kaikki oikeudet pidätetään.
© Begell House Inc. 2023
https://doi.org/10.1615/JPorMedia.2022043042
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi-fe2023021427199

Tiivistelmä

Determining pore sizes of porous media accurately is essential for many reasons. The standard ASTM F316 describing calculations used in gas-liquid displacement (GLD) porometry includes a factor B or a capillary constant. The standard does not explain the nature of this factor and the reasons for using its value. This work aims to investigate the capillary constant and to understand its meaning and applicability in calculating the pore diameter. Another goal is to assess the pore diameter measured in GLD porometry by comparing it with pore diameters measured with image analysis and cut point test methods. Seven filter media samples with as simple a structure as possible were analyzed. An investigation of the capillary constant revealed at least five different approaches, of which all are by no means applied for the calculations of pore size. Using the approach described in standard ASTM F316 to adjust the values obtained by GLD porometry was found unsuitable for the simply structured samples studied in this work. This work shows that the pore diameters measured by GLD porometry are most comparable with the other studied techniques when using Silwick as the wetting liquid. The raw data of bubble point pore and mean flow pore diameters measured by GLD porometry are similar to the largest pore sizes and mean pore diameters measured by image analysis and cut point test methods. This, in turn, confirms the absence of the need to use any constants for calculating the pore diameters for the samples studied in this work.

Lähdeviite

Ängeslevä, M., Salmimies, R., Rideal, G., Häkkinen A. (2023). Assessment of the Pore Diameter Measured in Gas-Liquid Displacement Porometry: Effect of the Capillary Constant. Journal of Porous Media, vol. 26, issue 1, pp. 31-49. DOI: 10.1615/JPorMedia.2022043042

Alkuperäinen verkko-osoite

https://www.dl.begellhouse.com/journals/49dcde6d4c0809db,37e3ce4f610e620a,1a3f968408661939.html
Kokoelmat
  • Tieteelliset julkaisut [1560]
LUT-yliopisto
PL 20
53851 Lappeenranta
Ota yhteyttä | Tietosuoja | Saavutettavuusseloste
 

 

Tämä kokoelma

JulkaisuajatTekijätNimekkeetKoulutusohjelmaAvainsanatSyöttöajatYhteisöt ja kokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy
LUT-yliopisto
PL 20
53851 Lappeenranta
Ota yhteyttä | Tietosuoja | Saavutettavuusseloste